5. Press the SRC-MAX PULSE RATE selection button. The SRC-MAX PULSE
RATE 200 LED lights. The monitor BPM increases and stabilizes to a value
between 197 and 203 inclusive. The following screen displays:
Active audio alarm.
Flashing %SpO
2
Flashing BPM indication between 197 and 203 inclusive.
Pulse waveform of approximately 1-inch P-T-P amplitude. Actual
amplitude may vary but will be a reference for low pulse amplitude/low
light patients.
6. Press the SRC-MAX PULSE RATE selection button. The SRC-MAX
PULSE RATE 60
stabilizes at a value between 57 and 63 inclusive. The following screen
displays:
Active audio alarm.
Flashing %SpO
2
BPM indication between 57 and 63 inclusive.
Pulse waveform of approximately 1-inch P-T-P amplitude. Actual
amplitude may vary, but can be used as a reference for low pulse
amplitude/low light patients.
%SP02
BPM
LIMITS
TREND
SETUP
LIGHT
indication between 73 and 77 inclusive.