Nellcor OxiMax N-600x Service Manual page 32

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

Performance Verification
26
Pulse Tone Volume Control
Adjust the pulse tone volume after completing the Alarm Volume Control on
page 25.
To set the pulse tone volume:
1. Press the ADJUST UP button and verify the sound level of the beeping pulse
tone volume increases.
2. Press the ADJUST DOWN button and verify the sound level of the beeping
pulse tone is silent.
3. Press the ADJUST UP button to return the beep volume to a comfortable
audible level.
4. Remove the O
M
XI
AX
from the monitor.
Nurse Call
To set the nurse call setting:
1. Connect the negative lead of a voltmeter to pin 5 and positive lead to pin 11 of
the Data Port Connector on the back of the monitor.
2. Ensure the audible alarm is not turned off.
3. Connect the SRC-MAX tester to the DOC-10 pulse oximetry cable.
4. Connect the DOC-10 cable to the monitor SpO
sensor from the finger and disconnect the DOC-10 cable
Data Port
Connector
Type BF
SN
2X
MADE IN IRELAND
T 0.50A 250V
CISPR 11
Group 1
Class B
100-
120
200-240V
R
50/60 Hz 30VA
0123
IPX1
C
US
N-600x
TYCO HEALTHCARE GROUP LP
NELLCOR PURITAN BENNETT INCORPORATED
PLEASANTON, CA USA 94588
AUTHORIZED REPRESENTATIVE:
TYCO HEALTHCARE UK LTD
GOSPORT PO13 0AS, UK
O
M
sensor port.
2
XI
AX