AX
Overview
The SRC-MAX functional tester enables qualified technicians to functionally test
Nellcor O
M
technology-based pulse oximeters and OEM O
XI
AX
technology-based monitors. The following table provides a brief description of
each test.
Tests
Descriptions
Test No. 1:
The test procedure simulates an O
BPM
sensor attached to a patient indicating 60 BPM
and 200 BPM.
Test No. 2:
The test procedure simulates an O
SpO
sensor attached to a patient indicating 75%
2
blood oxygen saturation and 90% blood
oxygen saturation.
Test No. 3:
The test procedure simulates an O
Modulation Level
sensor attached to a patient indicating low and
high pulse strength.
Test No. 4:
The test procedure simulates an O
Light
sensor attached to a patient indicating low and
high light level passing through the patient at
the sensor site.
Note: The SRC-MAX selectable indicator LEDs may extinguish if there is a
delay in proceeding through the above tests. This is normal operation in
order to increase the battery time.
Note: Pressing a button during the test procedures may be requested, changing a
certain parameter. If the LEDs are not lit, press the button twice. Pressing
the button once causes the indicators to relight and pressing twice initiates
the change.
Test No. 1: BPM
1. With the monitor turned off, connect the DOC-10 pulse oximetry cable to the
sensor port.
2. Connect the SRC-MAX tester to the other end of the DOC-10 cable.
Table 2:
Performance Verification
M
XI
AX
M
XI
AX
M
XI
AX
M
XI
AX
M
XI
AX