Power-On Settings and Service Functions
Exit Softkey
Next Softkey
52
The following list can be used as a quick reference showing how to reach different
softkey functions. Items reached through the PARAM softkey can be accessed
during normal operation. Functions provided by the PRINT and NEXT softkeys
cannot be accessed when a pulse oximetry cable is connected to the instrument.
Each of the various functions is described below.
PARAM (Service Menu)
-
RESET
-
-
RESET DEFAULTS?
-
-
-
YES (resets parameters to factory defaults, sounds three tones
indicating defaults are reset)
-
-
-
NO (back to Service menu)
-
SAVE
-
-
SAVE DEFAULTS?
-
-
-
YES (saves parameters as default settings, sounds three tones
indicating defaults are saved)
-
-
-
NO (back to Service menu)
-
SENSOR (enables/disables patient alarm events to the sensor memory chip)
-
-
BACK (back to Service menu)
-
-
EXIT (back to Main menu, sounds three tones indicating defaults are
reset)
-
BACK (back to Service menu)
PRINT
-
TREND
-
ERRLOG
-
INSTAT
-
INFO
NEXT
-
DOWNLD (for downloading monitor software)
-
ALARMS
-
-
SELECT
-
-
-
ALLOW OFF? (Allows alarms to be turned off) (up/down buttons
select Yes/No)
-
-
-
OFF REMINDER? (enables/disables Alarm Off reminder)
(up/down buttons select Yes/No)
-
-
-
ALLOW SAT-S? (enables/disables Sat-Seconds reminder)
(up/down buttons select Yes/No)
-
-
BACK (back to Service menu)
-
NEXT (back to Service menu)
-
-
BATTRY
-
-
NEXT
-
-
EXIT
(back to Main menu)
-
EXIT (back to Main menu)
The EXIT softkey returns the monitor to the Main menu.
The NEXT softkey enables you to view additional options available on various
menu levels.