Nellcor OxiMax N-600x Service Manual page 41

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

O
M
N-600x Service Manual
XI
AX
5. Press the SRC-MAX PULSE RATE selection button. The SRC-MAX PULSE
RATE 60 LED lights. The monitor BPM decreases and stabilizes to a value
between 67 to 73 BPM inclusive. The following screen displays.
Active audio alarm.
Flashing %SpO
2
BPM indication between 57 and 63 inclusive.
Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual
amplitude may vary but will be a reference for low pulse amplitude/low
light patients.
6. Turn off the monitor.
Test No. 2: SpO
2
1. With the monitor turned off, connect the DOC-10 pulse oximetry cable to the
sensor port.
2. Connect the SRC-MAX tester to the other end of the DOC-10 cable.
O
XI
%SP02
BPM
LIMITS
TREND
SETUP
LIGHT
indication between 73 and 77 inclusive.
Performance Verification
M
N-600x
AX
75
60
?
35