Nellcor OxiMax N-600x Service Manual page 83

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

O
M
N-600x Service Manual
XI
AX
Error Code
Description
531
External watchdog failure.
532
Power PC watchdog failure.
533
Boot NVROM uninitialized error.
534
Failed CRC check of application code in flash.
535
Failed periodic ram CRC check on application code running in
RAM.
560
Memory corruption detected.
561
RTOS Resource unavailable.
562
%SpO
front end reset.
2
563
%SpO
reported error.
2
564
Clinical mode was exited after input was received.
565
Communication failures between software modules.
566
Excessive resets before UIF runs.
567
An unexpected interrupt has been asserted.
568
General failure in UIF module generic post.
569
BOOT application program is corrupt - CRC does not match.
570
RTC was restarted.
574
Excessive restarts within one minute.
575
Trend data corrupted
576
Logic Processing Board (LPS) is going through excessive resets.
577
LPS status messages not arriving in a timely manner.
578
LPS driver has received too many corrupt (protocol error) messages.
579
Status report AC OK bit does not equal the hardware AC OK signal.
580
LPS driver received only a partial LPS message.
581
Too many corrupt (invalid data) message.
582
Unable to enter shelf-mode, this could mean that the LPS is dead or
there is a serial port transmit problem.
583
Primary speaker failure detected.
650
Oximetry module reports internal software consistency check failed.
651
Oximetry module reports software functions executed before
initialization completed.
652
Oximetry module reports internal memory buffer overflow.
653
Oximetry module reports host reset.
Table 12: Error Codes
Troubleshooting
77