Communication Protocol
O
M
N-600x Service Manual
XI
AX
4. Press the COMM softkey.
SERIAL PORT SETUP
BAUD
PROTOCOL
SELECT
5. Use the ADJUST UP and ADJUST DOWN buttons to select the desired baud
rate.
6. Press the EXIT softkey to set the baud rate. The baud rate setting will be in
effect until the monitor is powered off.
Note: The baud rate setup for the monitor may be saved as institutional default
settings. See WARNING: Audible alarms should not be silenced if patient
safety could be compromised. on page 60. Use care when performing this
procedure since all settings will be saved as institutional default settings.
The COMM softkey is used to select from eight communication protocols
supported by the data port. The selections are:
•
ASCII used for printouts
•
OXINET to enable communication with OXINET II
•
CLINICAL intended for Nellcor use only
•
GRAPH for graphic printouts
•
PHILIPS interfaces the N-600x with a Philips monitor
•
SPACELABS interfaces the N-600x with a SpaceLabs monitor
•
MARQ interfaces the N-600x with a GE Marquette monitor
•
DATEX interfaces the N-600x with a Datex-Ohmeda AS/3 monitor
Note: Selecting Philips, SpaceLabs, GE Marquette, or Datex-Ohmeda AS/3
automatically sets the baud rate to the rate applicable for that protocol.
%SP02
100.
9600
100
ASCII
BPM
BACK
EXIT
Data Port Interface Protocol