Nellcor OxiMax N-600x Service Manual page 126

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

Specifications
120
Table 18: Electromagnetic Immunity, RF Portable Equipment
The N-600x is suitable for use in the specified electromagnetic environment. The
customer and/or user of the N-600x should assure that it is used in an electromagnetic
environment as described below:
Immunity
IEC 60601-1-2
Test
Test Level
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz
800 MHz
3 V/m
800 MHz
2.5 GHz
Conducted RF
3 Vrms
IEC
150 kHz
61000-4-6
80 MHz
Note: Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast, and TV broadcast cannot be predicted theoretically with survey
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the N-600x is used exceeds the applicable
RF compliance level above, the N-600x should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the N-600x.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Interference may occur in the vicinity of equipment marked with the following symbol:
Electromagnetic
Compliance
Environment
Level
Guidance
3 V/m
distance = 1.2
80 MHz to 800 MHz
3 V/m
distance = 2.3
800 MHz to 2.5 GHz
3 Vrms
distance = 1.2
150 kHz to 80 MHz
Power
Power
Power