Next Softkey Menu - Nellcor OxiMax N-600x Service Manual

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

Power-On Settings and Service Functions

Next Softkey Menu

58
Additional options can be accessed from the Service Functions Main menu by
pressing the NEXT softkey. When NEXT is pressed, the softkeys change to the
functions shown below.
DOWNLD
ALARMS
DOWNLD Softkey
The DOWNLD softkey is used to display the revision of the Boot Code. To exit
DOWNLD, cycle power to the instrument by pressing the ON/STANDBY button.
Consult the Directions for Use (DFU) provided with any downloads or upgrades
to the FLASH firmware.
The baud rate for downloading new software is 19,200 via the data port.
ALARMS Softkey
The ALARMS softkey can change characteristics of the audible alarm. When the
ALARMS softkey is pressed, the softkey's functions change as shown below.
ALARMS
ALLOW OFF?
OFF REMINDER?
SELECT
O
M
N-600x
XI
AX
---
%SP02
---
BPM
NEXT
EXIT
?
N-600x
O
M
XI
AX
---
%SP02
YES
---
BPM
YES
BACK
?

Hide quick links: