Primary Speaker Failure

O
M
N-600x Service Manual
XI
AX
Table 13: Additional Display Messages
Message
Description
The
Sensor
oximetry cable, the cable has disconnected from the
Disconnected
monitor, or the
Press the ALARM SILENCE button to silence the alarm.
Check the connections. If this does not resolve the
problem, replace the
Settings Lost
Settings lost is displayed when the monitor detects the
current settings have been lost.
Unit Will Shut
The battery is critically low, and if the AC power is lost,
Down If AC Power
the monitor may power OFF.
Is Lost
The N-600x may detect a failure of the primary speaker and sound a high-pitched,
slow-pulsing piezo tone. A primary speaker failure message displays as shown
below:
PRIMARY SPKR FAILURE:
NOTIFY SVC PERSONNEL.
PRESS HELP
HELP
1. Press HELP to continue. The following message displays.
HELP SPEAKER FAILURE
Note: Once this
monitor is powered
off, it cannot be
powered on again.
2. Press BACK to display the speaker failure message again. The message
cannot be cleared.
O
M
sensor has disconnected from the pulse
XI
AX
O
M
sensor/cable wiring is defective.
XI
AX
O
M
sensor and/or cable.
XI
AX
100
100
BACK
Troubleshooting