O
M
N-600x Service Manual
XI
AX
Error Code
Description
721
Battery charger error - instance #2.
722
Battery charger error - instance #3.
723
Battery charger error - instance #4.
724
Battery charger error - instance #5.
725
Battery charger error - instance #6.
726
Battery charger error - instance #7.
727
Battery charger error - instance #8.
728
Battery charger error - instance #9.
729
Battery critical without AC power.
730
Battery fail without AC power.
731
EEPROM CRC Error - EEPROM cleared.
732
Vref error on DC.
733
Vref error on AC.
734
Overtemperature (> 58 C).
735
Battery I charge.
736
Battery I discharge.
737
Temperature low.
738
Hardware failsafe occurred, maybe overtemperature.
739
Hardware failsafe occurred, maybe overcurrent.
740
Hardware failsafe occurred, cause unknown.
741
Message RX CRC error.
743
Error in the shut down sequence - occurs if the power is cycled off
and on too quickly.
Table 12: Error Codes