Setting Institutional Defaults (Sample)

N-595
6. Use the ADJUST UP or ADJUST DOWN button to change selected
parameter.
7. Press the BACK softkey.
When ALLOW OFF is selected, a choice is given between allowing an audible
Alarm Off or disabling the audible alarm OFF. Pressing the ADJUST UP or
ADJUST DOWN button cycles between YES and NO. If YES is selected, the
operator has the option of selecting AUDIBLE ALARM OFF. If NO is selected,
the operator is not given the option of selecting AUDIBLE ALARM OFF as an
alarm silence duration choice.
If the audible alarm is set to Off, a reminder tone can be sounded every 3 minutes
to notify the user of this condition. The ADJUST UP and ADJUST DOWN
buttons can be used to change the choice from YES to NO. Selecting YES enables
the Reminder. Selecting NO disables the Reminder when the audible alarm is set
to Off.
Power-up default values may be changed to institutional power-up default values.
Set the desired limits in the normal operation mode and you will set the
institutional defaults in the monitor's service mode. The following default values
may be set:
Alarm Silence Duration (30, 60, 90, 120 seconds)
Alarms (Allow Off - Yes/No)
Off Reminder (Yes/No)
Alarm Volume (1 to 10)
Sensor Event Record Type (SpO
Nurse Call Priority RS-232 (normally high, normally low)
Pulse Beep Volume (0 to 10)
Pulse Rate Upper Alarm Limit (lower limit plus 1 to 250 bpm)
Pulse Rate Lower Alarm Limit (20 bpm to upper limit minus 1)
SatSeconds (OFF, 10, 25, 50, 100)
Sensor Adjust Enable (Yes/No)
Data Port Baud Rate (2400, 9600, 19200)
Power-On Settings and Service Functions
, SpO
+ BPM, Default)
2
2
47