Nellcor OxiMax N-600x Service Manual page 61

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

O
M
N-600x Service Manual
XI
AX
Trend Softkey
A Trend printout includes all data recorded for up to 48 hours of monitoring since
the last Delete Trends was performed. A new trend point is recorded every four
seconds. The figure below is an example of a Trend printout.
N-600x VERSION X.X.X.X
ADULT
0SAT-S
TIME
01-JAN-06 14:00:00
01-JAN-06 14:00:05
01-JAN-06 14:00:10
01-JAN-06 14:00:15
01-JAN-06 18:00:43
01-JAN-06 18:00:48
N-600x VERSION X.X.X.X
ADULT
0SAT-S
TIME
01-JAN-06
18:00:53
01-JAN-06
18:00:58
01-JAN-06
18:01:03
01-JAN-06
18:01:08
01-JAN-06
18:01:13
Output Complete
The first row of the printout includes information about the type of instrument
delivering the information, the software level, type of printout, and alarm
parameters. The second line lists the headings for the columns. These lines are
printed out every 25 lines, or when a change to an alarm limit is made.
Patient data is represented with a date and time stamp for the data. In the example
above, the [ - - - ] means that an O
quality of the data being received was too low for the monitor to interpret the data.
Patient data that is outside of an alarm limit is marked with an asterisk (*).
At the end of the printout "Output Complete" prints indicating the data is not
corrupt. If the Output Complete statement is not printed at the end of the printout,
the data must be considered invalid.
Power-On Settings and Service Functions
TREND
SpO2 Limit: 30-100%
SPO2 RESP MODE: NORMAL
%SpO2
PR (bpm) PA
100
120
220
100
124
220
100
190
220
100
190
220
- - -
- - -
- - -
- - -
- - -
- - -
TREND
SpO2 Limit: 30-100%
SPO2 RESP MODE: NORMAL
%SpO2
PR (bpm) PA
- - -
- - -
- - -
- - -
- - -
- - -
98
100
140
98
181*
190
99
122
232
M
sensor was connected but the signal
XI
AX
PR Limit: 100-180BPM
PR Limit: 100-180BPM
55