Nellcor OxiMax N-600x Service Manual page 29

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

O
M
N-600x Service Manual
XI
AX
Alarms and Alarm Silence
SpO
O
M
XI
AX
2
Sensor Port
LIMITS
To adjust the alarms and alarm silence options:
1. Connect the DOC-10 pulse oximetry cable to the monitor's SpO
sensor port.
2. Connect the DS-100A O
3. Press the ON/STANDBY button to turn the monitor on.
4. Press SETUP and then VIEW.
5. Press the Pleth softkey. Verify the %SpO
pulse rate.
6. Press the LIMITS softkey.
7. Press the SELECT softkey to select %SpO
8. Press the ADJUST UP button until the %SpO
99.
9. Press the SELECT softkey three consecutive times to select Pulse Rate Lower
Alarm Limit.
10. Press the ADJUST UP button until the Pulse Rate Lower Alarm Limit
indicates 160.