Performance Verification
20
5. Press the SELECT softkey. Verify the monitor emits a single beep and the
%SpO
Lower Alarm Limit of "85" (or your institutional default setting) is
2
selected.
ADULT LIMITS
UPPER
LOWER
SAT-S
SELECT
6. Press and hold the ADJUST DOWN button. Verify the %SpO
Limit display reduces to a minimum of "20."
7. Press and hold the ADJUST UP button. Verify the %SpO
Limit display cannot be raised past the Upper Alarm Limit setting of "85."
8. Press the LIMITS softkey.
9. Press the SELECT softkey three consecutive times. Verify the monitor emits a
beep after each keystroke. The selected Pulse Upper Alarm Limit should be
"170."
ADULT LIMITS
UPPER
LOWER
SAT-S
SELECT
10. Press and hold the ADJUST DOWN button. Verify the minimum displayed
value is "41" for the BPM Upper Alarm Limit.
11. Press the LIMITS softkey.
O
%SP02
%SPO2
BPM
86
170
85
40
BPM
OFF
NEO
ADULT
EXIT
O
%SP02
%SPO2
BPM
170
86
40
BPM
85
OFF
NEO
ADULT
EXIT
M
XI
AX
N-600x
100
110
?
Lower Alarm
2
Lower Alarm
2
M
N-600x
XI
AX
100
110
?