Patient Applied Risk Current; Patient Isolation Risk Current - Nellcor OxiMax N-600x Service Manual

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

Patient Applied Risk Current

Patient Isolation Risk Current

O
M
N-600x Service Manual
XI
AX
The following table displays the maximum patient applied risk current for a BF
type applied part.
Neutral
AC Line Polarity
Line
Normal
Closed
Normal
Open
Normal
Closed
Reversed
Closed
Reversed
Open
Reversed
Closed
The following table displays the maximum patient isolation risk current for a BF
type applied part.
Neutral
AC Line Polarity
Line
Normal
Closed
Reversed
Closed
Power Line
IEC 60601-1
Ground Cable
UL 60601-1
Closed
100 µA
500 µA
Closed
500 µA
Open
Closed
100 µA
Closed
500 µA
500 µA
Open
Power Line
IEC 60601-1
Ground Cable
UL 2601-1
Closed
5 mA
Closed
5 mA
Specifications