Performance Verification
36
3. Turn on the monitor by pressing the ON/STANDBY button. After the monitor
completes it self-check, the following screen displays:
•
Active audio alarm
•
Flashing %SpO
2
•
BPM indication between 57 and 63 inclusive
•
Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual
amplitude may vary but will be a reference for low pulse amplitude/low
light patients
4. Press the SRC-MAX %SpO
LED lights.
The monitor displays three dashes [ - - - ] until the %SpO
value between 88 and 92 inclusive. The following screen displays:
•
No audio alarm.
•
%SpO
indication between 88 and 92 inclusive.
2
•
BPM indication between 57 and 63 inclusive.
•
Pulse waveform of approximately 1/2-inch P-T-P amplitude. Actual
amplitude may vary but will be a reference for low pulse amplitude/low
light patients.
O
%SP02
BPM
LIMITS
TREND
SETUP
LIGHT
indication between 73 and 77 inclusive
selection button. The SRC-MAX %SpO
2
%SP02
LIMITS
TREND
SETUP
LIGHT
M
N-600x
XI
AX
75
60
?
stabilizes at a
2
O
M
N-600x
XI
AX
90
60
BPM
?
90
2