P o w e r - O n S e t t i n g s a n d S e r v i c e F u n c t i o n s

Overview

Adjusting the Power-On Settings

O
M
N-600x Service Manual
XI
AX
This chapter discusses how to reconfigure power-on default values and access the
service functions.
By using the softkeys as shown in Figure 1, you can the change alarm limits,
displays, baud rates, time and date, and trends to view.
Some values cannot be saved as power-on default values. An %SpO
Alarm Limit less than 85 will not be saved as a power-on default. AUDIBLE
ALARM OFF will not be accepted as a power-on default. An attempt to save
either of these values as default results in an invalid tone. These limits can be
adjusted lower for the current patient, but erased when the instrument is turned
off.
A decimal point is added to the right of a display when the alarm limit for that
display has been changed to a value that is not a power-on default value. If the
new value is saved as a power-on default value, the decimal point is removed. By
using the service functions, changes can be saved as power-on default values.

Power-On Settings and Service Functions