Nellcor OxiMax N-600x Service Manual page 27

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

O
M
N-600x Service Manual
XI
AX
12. Press the SELECT softkey four consecutive times. Verify the selected Pulse
Rate Lower Alarm Limit display indicates an alarm limit of "40."
ADULT LIMITS
UPPER
LOWER
SAT-S
SELECT
13. Press and hold the ADJUST DOWN button. Verify the selected Pulse Rate
Lower Alarm Limit display decreases to a minimum of "30."
14. Press and hold the ADJUST UP button. Verify the selected Pulse Rate Lower
Alarm Limit display cannot be adjusted above the Pulse Rate Upper Alarm
Limit of "40."
15. Press the LIMITS softkey.
16. Press the SELECT softkey twice. Verify the SatSeconds SAT-S alarm is
selected.
ADULT LIMITS
UPPER
LOWER
SAT-S
SELECT
17. Press the ADJUST UP button repeatedly until the SatSeconds alarm display
cycles from OFF through 10, 25, 50, 100, and OFF.
18. Press the ON/STANDBY button to turn the monitor off.
19. Press the ON/STANDBY button to turn the monitor back on.
20. Press the LIMITS softkey. Verify the selected %SpO
display indicates an alarm limit of "100."
O
%SP02
BPM
%SPO2
41
86
BPM
85
40
OFF
NEO
ADULT
EXIT
O
%SP02
BPM
%SPO2
41
86
BPM
85
40
OFF
NEO
ADULT
EXIT
Performance Verification
M
N-600x
XI
AX
100
110
?
M
XI
AX
N-600x
100
110
?
Upper Alarm Limit
2
21