Electrical Requirements - Nellcor OxiMax N-600x Service Manual

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

Specifications

Electrical Requirements

112
1
Accuracy
For a complete listing of SpO
sensors, see the Sensor Accuracy Grid online at: http://www.nellcor.com, or contact
1-800-NELLCOR.
1
Subjects used to validate SpO
the local population. Comprised of both men and women, subjects spanned a range of
skin pigmentations and ranged in age from 18-50 years old. Accuracy specifications are
based on controlled hypoxia studies with healthy non-smoking adult volunteers over the
specified saturation SpO
2
SaO
values of drawn blood samples measured by hemoximetry. All accuracies are
2
expressed as ± "X" digits. Pulse oximeter equipment measurements are statistically
distributed; about two-thirds of pulse oximeter measurements can be expected to fall in
this accuracy (A
) range. Because scatter and bias of pulse oximeter SpO
RMS
SaO
comparison commonly increase as the saturation decreases, and accuracy
2
specifications are calculated from data spanning the stated range, different accuracy
values may result when described partially overlapping ranges.
2
Clinical functionality has been demonstrated on a population of hospitalized neonate
patients. The observed SpO
to 23 days, weight from 750 to 4,100 grams, and 63 observations made spanning a range
of 85% to 99% SaO
while monitored with the Nellcor
2
3
Specification applies to N-600x monitor performance. Reading accuracy in the presence
of low perfusion (detected IR pulse modulation amplitude 0.03% - 1.5%) was validated
using signals supplied by a patient simulator. SpO
across the monitoring range over a range of weak signal conditions and compared to the
known true saturation and pulse rate of the input signals.
Display Update Interval
2 seconds
Instrument
Power Requirements
Fuses
Battery
The battery provides at least 7 hours of battery life when new and fully-charged with no
alarms, no serial data, no analog output, no nurse call output, with backlight on while
using a pulse simulator set for 200 bpm, high light and low modulation.
Type
accuracy across the full line of available Nellcor
2
measurement accuracies were healthy and recruited from
2
range(s). Pulse oximeter SpO
2
accuracy was 2.5% in a study of 42 patients with ages of 1
2
O
and pulse rate values were varied
2
Rated at 100 to 120 volts AC (nominal 120 VAC) or 220
to 240 volts AC (nominal 230 VAC), 30 volt/amps to be
compliant with IEC 60601-1 sub-clause 10.2.2
Quantity: 2, 0.5 A, 250 volts, slow-blow,
IEC (5 x 20 mm)
Lead acid
O
M
XI
AX
readings were compared to
and blood
2
M
N-595 pulse oximeter.
XI
AX