Nellcor OxiMax N-600x Service Manual page 136

Pulse oximeter
Hide thumbs Also See for OxiMax N-600x:
Table of Contents

Advertisement

Data Port Interface Protocol
130
To change the communication protocol:
1. Turn on the monitor by pressing the ON/STANDBY button.
2. Press the SETUP softkey.
3. Press the NEXT softkey twice.
4. Press the COMM softkey.
5. Press the SELECT softkey.
6. Use the ADJUST UP and ADJUST DOWN buttons to select the desired
protocol.
7. Press the EXIT softkey set the protocol. The protocol setting is in effect until
the monitor is powered off.
Note: The protocol setup for the monitor may be saved as institutional default
settings. See WARNING: Audible alarms should not be silenced if patient
safety could be compromised. on page 60. Use care when performing this
procedure since all settings will be saved as institutional default settings.