Column Heading

O
M
N-600x Service Manual
XI
AX
The first two lines of the chart are the column headings shown below. Every 25th
line a column heading is printed. A column heading is also printed when a value
of the column heading is changed. There are three column headings shown in
Figure 17. The third column heading was printed since the %SpO
from 85-100% to 80-100%.
N-600x VERSION X.X.X.X CRC: XXXX SpO2 Limit: 70-100%
ADULT
TIME
Data Source
N-600x VERSION X.X.X.X CRC: XXXX SpO2 Limit: 70-100%
ADULT
TIME
Data in the highlighted box above represents the source of the printout or display,
in this case the monitor.
Software Revision Level
N-600x VERSION X.X.X.X CRC: XXXX SpO2 Limit: 70-100%
ADULT
TIME
The next data field tells the user the software level (Version X.X.X.X) and a
software verification number (CRC XXXX). Neither of these numbers should
change during normal operation. The numbers change if the monitor is serviced
and receives a software upgrade.
Alarm Limits
N-600x VERSION X.X.X.X CRC: XXXX SpO2 Limit: 70-100%
ADULT
TIME
The last data field in the top line indicates the upper and the lower alarm limits for
%SpO
and for the pulse rate (PR). In the example above, the lower alarm limit
2
for %SpO
is 70% and the upper alarm limit is 100%. Pulse Rate alarm limits are
2
60 BPM (lower), and 160 BPM (upper).
0SAT-S
SPO2 RESP MODE: NORMAL
%SpO2
BPM
PA
Status
0SAT-S
SPO2 RESP MODE: NORMAL
%SpO2
BPM
PA
Status
0SAT-S
SPO2 RESP MODE: NORMAL
%SpO2
BPM
PA
Status
0SAT-S
SPO2 RESP MODE: NORMAL
%SpO2
BPM
PA
Status
Data Port Interface Protocol
limits changed
2
PR Limit: 60-160BPM
PR Limit: 60-160BPM
PR Limit: 60-160BPM
PR Limit: 60-160BPM
141