Performance Verification; Introduction; Equipment Needed; Table 1. Required Test Equipments - Nellcor OxiMax N 5500 Service Manual

Patient monitor
Table of Contents

Advertisement

Section 3: Performance Verification

3.1 Introduction

3.2 Equipment Needed

3.3 Performance Tests
3.4 Safety Tests
3.1 Introduction
This section discusses the tests used to verify performance following repairs or during
routine maintenance. All tests can be performed without removing the N5500 covers.
All tests except the battery charge and battery performance tests must be performed as
the last operation before the monitor is returned to the user.
If the N5500 fails to perform as specified in any test, repairs must be made to correct
the problem before the monitor is returned to the user.
3.2 Equipment Needed
Table 1 lists the equipment required for performance verification.
Equipment
Digital multimeter (DMM)
ECG cable for 3-leadwires
ECG cable for 5-leadwires (optional)
ECG leads
ECG leads (optional)
NIBP hose
NIBP cuff
Pulse oximetry cable
O
M
XI
AX
O
M
XI
AX
Temperature probes
SpO
simulator
2
ECG simulator
NIBP simulator
Respiration simulator
Temperature simulator
Safety analyzer
Data interface cable
Stopwatch

Table 1. Required Test Equipments

Durasensor® oxygen transducer
oxygen transducer
N5500 Service Manual
Description
Fluke Model 87 or equivalent
MDEC No.2
MDEC No.4
MDEL series (MDEL-3S, MDEL-3G)
MDEL series (MDEL-5S, MDEL-5G)
Colin Hose No.1EM/No.3EM
Colin Cuff series
DOC-10
DS-100A
MAX-A
Monotherm™
Nellcor SRC-MAX simulator
METRON PS-420 or equivalent
Bio-Tek "BP Pump 2" or equivalent
METRON PS-420 or equivalent
medSim 300 or equivalent
METRON QA-90 or equivalent
RS-232 cable
Manual or electronic
9