Performance Verification; Introduction; Equipment Required; Figure 10: Pinout Reference Src-Max - Nellcor OxiMax SRC-MAX Technical Manual

Pulse oximetry functional tester
Table of Contents

Advertisement

P e r f o r m a n c e Ve r i f i c a t i o n
Note:

Introduction

This section describes the procedures for verifying specified
functionality of the SRC-MAX.
All measurements/tests shall be performed through the 9-pin
sensor/cable connector. See Figure 10.
Caution: Do not open the case of the SRC-MAX. The
device could be damaged.
Caution: Do not connect ground to any pins on the SRC-
MAX 9 pin connector or the 9-pin D-sub female connector
used in the test set up. The device could be damaged.
Refer to Block Diagram, Figure 11 on page 22.

Equipment Required

SRC-MAX
In this manual, pulse oximeters refer to Nellcor
O
M
technology-based pulse oximeters and
XI
AX
monitors refer to OEM O
monitors.

Figure 10: Pinout Reference SRC-MAX

Oscilloscope; 20 MHz bandwidth, 2 channels, able to
resolve 500 millivolts peak-to-peak to 3 volts peak-
to-peak at 1,000 Hz, 1 and 3 Hz
10 ohm ±1% resistor; 1/10 Watt Metal Film
1 meg-ohm ±1% resistor; 1/10 Watt Metal Film
Power Supply, able to deliver +5 volts 100 mA
M
technology-based
XI
AX
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents