YOKOGAWA GX10 User Manual page 41

Gx series; gp series; gm series; smartdac+ communication command
Hide thumbs Also See for GX10:
Table of Contents

Advertisement

You cannot use this command to configure settings
while computation is in progress.
If p2=TC/RTD/DI/OHM, you cannot specify p4=Sqrt.
If p2=GS, you cannot specify p4=Off/Delta.
If p2=DI, you cannot set p7.
If an electro-magnetic relay type or low withstand
voltage relay type analog input module is in use, you
cannot specify p2=RTD.
For 4-wire RTD modules, p2 cannot be set to Volt, TC,
GS, or DI. (Only SKIP, RTD, and OHM are selectable.)
p2 = OHM is valid only for 4-wire RTD modules.
The settable items for p3 are shown below.
p2=Volt p2=TC
p2=RTD
20mV
R
Pt100
60mV
S
Pt100-H
200mV
B
JPt100
1V
K
JPt100-H
2V
K-H
Cu10GE
6V
E
Cu10LN
20V
J
Cu10WEED
50V
T
Cu10BAILEY
1
100V
N
Cu10a392
W
Cu10a393
L
Cu25
U
Cu53
PLATINEL
Cu100
PR20-40
J263B
WRe3-25
Ni100SAMA
KpvsAu7Fe
Ni100DIN
NiNiMo
Ni120
WWRe26
Pt25
N14
Pt50
XK
Pt200WEED
Cu10G
Cu50G
Cu100G
Pt46G
Pt100G
Pt500
Pt1000
1 Valid only for high-speed AI modules
2 Valid only for 4-wire RTD modules
If p4=LogT1 on a Log scale channel, set the value in
the following range.
p5<p6
p9, p10
p9<p10. The maximum span is 15 decades.
If the mantissa of p9 is 1.00, the minimum span is
1 decade.
If the mantissa of p9 is not 1.00, the minimum
span is 2 decades.
If p4=LogT2 or LogT3 on a Log scale channel, set the
value in the following range.
p5<p6
p9, p10
The maximum span is 15 decades; the minimum
is 1 decade.
If the mantissa of p9 is not 1.00, the exponent
is +14 or less, and the maximum span is 14
decades.
IM 04L51B01-17EN
p2=GS
p2=DI
1-5V
Level
0.4-2V
DI
2
2
SRangeAICurrent
Measurement Range of Current Input
Type AI Channel
Sets the measurement range of an current input type AI
channel.
Unused Channels
Syntax
SRangeAICurrent,p1,p2
p1 Channel number
p2 Input type (Skip)
Channels Whose Input Type is Current and No Math
Syntax
SRangeAICurrent,p1,p2,p3,p4,p5,p6
,p7
p1 Channel number
p2 Input type (Current)
p3 Range (0-20mA)
p4 Math type (Off)
p5 Span lower limit
p6 Span upper limit
p7 Bias (–999999 to 999999)
Delta Channels
Syntax
SRangeAICurrent,p1,p2,p3,p4,p5,p6,
p7,p8
p1 Channel number
p2 Input type (Current)
p3 Range (0-20mA)
P4 Math type (Delta)
p5 Span lower limit
p6 Span upper limit
p7 Bias (–999999 to 999999)
p8 Reference channel number
Scaling Channels
Syntax
SRangeAICurrent,p1,p2,p3,p4,p5,p6,
p7,p8,p9,p10,p11
p1 Channel number
p2 Input type (Current)
p3 Range (0-20mA)
P4 Math type (Scale)
p5 Span lower limit
p6 Span upper limit
p7 Bias (–999999 to 999999)
p8 Decimal place (0 to 5)
p9 Scaling lower limit
p10 Scaling upper limit
p11 Unit (up to 6 characters, UTF-8)
Scaling Channels (General Signal 4-20 mA Input)
Syntax
SRangeAICurrent,p1,p2,p3,p4,p5,p6,
p7,p8,p9,p10,p11,p12,p13
p1 Channel number
p2 Input type (GS)
p3 Range (4-20mA)
P4 Math type (Scale)
2.4 Setting Commands
2-23
1
2
App

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gm10Gx20Gp20Gp10

Table of Contents