YOKOGAWA GX10 User Manual page 120

Gx series; gp series; gm series; smartdac+ communication command
Hide thumbs Also See for GX10:
Table of Contents

Advertisement

2.6 Operation Commands
All
1
2
Example When the measurement mode is dual interval,
apply a trigger to scan group 2.
OExecRec,1,2
Description
Manual trigger (p1 = 1) cannot be executed when the
advanced security function (/AS) is enabled.
If a manual sample is executed (p1 = 0) when there
are no source channels for manual sampling, a file
without any source channels will be created.
p2 is valid when multi batch is enabled and p1=3 or 4.
p2 is valid when the measurement mode is set to dual
interval and p1=1.
OExecSNTP
Queries the Time Using SNTP
Queries the time using SNTP.
Syntax
OExecSNTP,p1
p1 Time query execution (0)
Example Query the time using SNTP.
OExecSNTP,0
OMessage
Message Writing
Writes a message.
Write a Preset Message
Syntax
OMessage,p1,p2,p3,p4
p1 Action type (PRESET)
p2 Message number (1 to 100)
p3 Display group number
ALL
1 to 60 Write to specified groups
You can specify multiple groups at once.
To do so, separate display groups with a
colon.
p4 Batch group number (1 to the number
used)
Example Write the message in preset message number
8 to display groups 1 and 2.
OMessage,PRESET,8,1:2
Description
p4 is valid when multi batch is enabled. This cannot
be omitted.
Write a Free Message
Syntax
OMessage,p1,p2,p3,p4,p5
p1 Action type (FREE)
p2 Message number (1 to 10)
p3 Display group number
ALL
1 to 60 Write to specified groups
2-102
All scan groups
Scan group 1
Scan group 2
Write to all display groups
Write to all display groups
You can specify multiple groups at once.
To do so, separate display groups with a
colon.
P4 Message string to write (up to 32
characters, UTF-8)
p5 Batch group number (1 to the number
used)
Example Write a free message "MARK" as message
number 2 in display groups 3, 8, and 11.
OMessage,FREE,2,3:8:11,'MARK'
Description
p5 is valid when multi batch is enabled. This cannot
be omitted.
OPassword
Changes the Password
Changes the password.
Syntax
OPassword,p1,p2,p3
p1 Old password (up to 20 characters, ASCII)
p2 New password (up to 20 characters,
ASCII)
p3 New password (enter the same password
as p2)
Example Change the password from "PASS001" to
"WORD005."
OPassword,'PASS001','WORD005','WO
RD005'
Description
For the characters that you can use for the password,
Appendix
1.
see
OMath
Starts, Stops, or Resets Computation or
Clears the Computation Dropout Status
Display
Starts or stops computation, resets computed values, or
clears the computation dropout status display.
Syntax
OMath,p1,p2
p1 Action type (0)
0
1
2
3
p2 Batch group number
All
1 to the
number
used
Query
OMath?
Example Start computation.
OMath,0
Description
You cannot use this command while the recorder is
saving or loading setup data.
Start computation
Stop computation
Reset computation
Clear the computation dropout
status display
All math channels
Math channel belonging to the
specified batch group
IM 04L51B01-17EN

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gm10Gx20Gp20Gp10

Table of Contents