General Communication Log (Flog); Im 04L51B01-17En - YOKOGAWA GX10 User Manual

Gx series; gp series; gm series; smartdac+ communication command
Hide thumbs Also See for GX10:
Table of Contents

Advertisement

2.10 ASCII Output Format

2.10.14 General Communication Log (FLog)

2-138
The output in response to the command "FLog,General" is shown below.
Syntax
EA<crlf>
yyyy/mo/dd_hh:mm:ss_nn_uuu...u_fdmmm...m<crlf>
...
EN<crlf>
yyyy/mo/dd_hh:mm:ss Time of command Tx/Rx
nn
uuu...u
f
d
mmm...m
yyyy
mo
dd
hh
mm
ss
Connection ID
s0
s1
s2
e0
e1
e2
e3
User name (fixed to 20 characters. Unused character positions
are filled with spaces.)
Multiple command flag
Space
*
Tx/Rx
>
<
Message (fixed to 40 characters. Unused character positions are
filled with spaces.)
The recorder normally outputs the data that has been transmitted
or received as-is, but it sometimes outputs special messages.
Special messages are shown below.
(output)
(Over length)
(timed out)
(disconnected)
Year (1900 to 2099)
Month (01 to 12)
Day (01 to 31)
Hour (00 to 23)
Minute (00 to 59)
Second (00 to 59)
Serial (general)
Bluetooth connection
USB connection
Ethernet connection #0 (general)
Ethernet connection #1 (general)
Ethernet connection #2 (general)
Ethernet connection #3 (general)
Single command
Multiple commands
Tx (command: connected instrument to
recorder)
Rx (Response: recorder to connected
instrument)
Data output
Command length too long
Timeout
Disconnection (occurs when an
Ethernet connection is disconnected)

IM 04L51B01-17EN