Replace The Peristaltic Pump Cartridge - Agilent Technologies 1260 Infinity II User Manual

Infinitylab lc series preparative autosampler
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

Replace the Peristaltic Pump Cartridge

When
Tubing blocked or broken
Tools required
Description
Parts required
p/n
5065-4445
Preparations
Remove the inlet filter of the solvent bottle which guides the solvent to the peristaltic pump to
avoid syphoning effects.
When opening capillary or tube fittings solvents may leak out.
WA R N I N G
The handling of toxic and hazardous solvents and reagents can hold health risks.
The peristaltic pump cartridge is a replaceable unit. The tubing inside the pump is not
N O T E
replaceable.
Agilent 1260 Infinity II Preparative Autosampler User Manual
Description
Peristaltic pump with Pharmed tubing
Please observe appropriate safety procedures (for example, goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are
used.

Replace the Peristaltic Pump Cartridge

9
Maintenance
179

Advertisement

Table of Contents
loading

Table of Contents