8
Error Information

Sample Cooler Error Messages

Sample Cooler Error Messages
Cooler Voltage Failure
Error ID: 30722
The sample cooler board has detected a voltage failure.
Probable cause
1
2
Cooler PCB Error State
Error ID: 30724, 30275
The sample cooler PCB board seems to be defective.
Probable cause
1
2
144
Blown fuses, defective cable
Voltage issue in the sample cooler
Blown fuses, defective cable
Defective PCB boards
Suggested actions
• Check the power cable from the ALS to the
sample cooler.
• Check if a fuse is blown (LED status on the
specific mainboard).
Replace the sample cooler first then check the
specific mainboard.
Suggested actions
• Check the power cable from the ALS to the
sample cooler.
• Check if a fuse is blown (LED status on the
specific mainboard).
Replace the sample cooler.