Transport And Installation - h/p/cosmos Arsalis Gaitway 3D Service Manual

Table of Contents

Advertisement

gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
 Operate the device at a temperature of +10 °C to +40 °C with relative humidity of
30 % to 70 % non-condensing, air pressure of 700 to 1060 hPa (maximum 3000m
altitude).
More information on room conditions for the gaitway-3D treadmill is
available in the instructions for use of the h/p/cosmos treadmill
ergometer. Please refer to the manual available at the address below:
http://www.h-p-cosmos.com/en/company/downloads.htm
3.3.

Transport and installation

 The device must be installed by trained and qualified technicians certified by
Arsalis.
 After the installation, the device must remain bolted to the floor for safe and
optimal operation.
 Any transport of the device requires supervision by trained and qualified
technicians certified by Arsalis.
 The manufacturer does not assume liability for any damage, complaints or
missing parts which are not reported immediately upon delivery on the packing
list/delivery note.
 Failure to comply with the conditions listed in this service manual and listed in the
operation and service manuals of other respective devices which may be used in
connection with this device, failure of performing recommended maintenance and
safety inspection intervals, unauthorized maintenance or amendments of the
design and/or performance and/or specifications and/or labelling of the devices
shall absolve Arsalis from any responsibility for the safety, reliability and
performance of this equipment.
 Always ground (earth) the device to prevent electric shock. If the power supply
outlet is not grounded, it will be necessary to install a ground by qualified
engineers.
 It should be ensured that all cables (power connection, Ethernet, interface,
potential equalization, etc.) and accessories are installed properly and safely and
that nobody can stumble or fall over the cables and/or accessories.
 All parts should be kept away from children and animals.
 Only connect accessories, software and host equipment if it is confirmed as
compatible by all manufacturers.
 It is not allowed to connect sports devices to medical devices such as ECG-
systems.
 Connecting a running machine to a medical device results in a medical system.
Only trained staff is authorized to perform this connection. Always use IEC 60601-
1 approved potential isolation components. This medical system is to be
connected via a potential balance cable with the provided connector bolt and
bearing within the designated room.
 Do not touch any electrical parts (such as power supply plug) with a wet hand.
This may cause electric shock.
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 9