Figure 2-1
Four-wire sense connections for battery and charger channels
DVM Input
DVM +
DVM -
Source -
Source -
Sense -
Sense +
Source +
Source +
Model 2306
Source Input/Output

Remote sense

As shown
remote sense leads (4 wire connection). The Sense+ and Sense- pins provide output voltage
sensing. Without these terminals connected, the power supply operates without voltage feedback
and therefore supplies an unregulated voltage. This unregulated voltage value can be up to +18V
or down to -5V. Use voltage protection to turn off the output and protect against the extremes
(refer to "Setting voltage protection value" on page 2-5).
Connect the sense inputs to the supply as close as possible to the load's source inputs through
twisted pair leads (refer to
formance of the supply.
NOTE
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Quick Disconnect
Connector
Twisted Pair
inFigure 2-1
the 2306 battery and charger channels are intended to be operated with
Figure
Do NOT jumper the sense inputs and supply outputs at the rear of the supply! Con-
necting the sense leads in this fashion will severely compromise the performance of
Model 2306 with dynamic loads when using 4-wire sense.
Basic Power Supply Operation
2-1). This is necessary to achieve the maximum transient per-
External Test Circuit
+
-
DUT
2-3

Hide quick links: