Figure 3-5
Sample pulse forms for step method
4 Up
3 Down
Figure 3-6
Sample one-shot only pulses for step method
Trigger level settings
The trigger level may be set to a unique value for each active step. Use the TLEVx command
to set appropriate trigger levels for each active step in the waveform. Make sure that the maxi-
mum setting for the selected trigger level range is not exceeded.
page
3-29.)
Figure 3-7
has 5 rising edge steps and 4 falling edge steps. Set the trigger levels for each step
measurement according to the expected pulses. Based on the wave form, the nine trigger levels
could be set as follows:
Rising:
TLEV1
TLEV2
TLEV3
TLEV4
TLEV5
For a programming example of this sample, see
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
0 Up
4 Down
0 Up
5 Down
Falling:
100mA
TLEV6
300mA
TLEV7
500mA
TLEV8
700mA
TLEV9
900mA
"Sample step method" on page
Pulse Current Measurements
5 Up
0 Down
5 Up
0 Down
(See "Trigger level range" on
900mA
600mA
400mA
300mA
3-32.