Remote display option

NOTE
If mounting the power supply in a location where the display cannot be seen or the controls
are not easily accessible, use the optional Model 2304-DISP or 2306-DISP Display Module (see
Figure
(with the exception of power). All features/menus work as described for the Model 2306
(exceptions are noted). A 9 foot cable attaches the remote display to the rear of the power supply
allowing the unit to be operated remotely.
Figure 1-4
2304-DISP Remote display
option (2306-DISP similar)
NOTE
Plug the remote display module into the rear panel connector labeled "REMOTE DISPLAY
OPTION" (see rear panel in
with the following message displayed:
When the remote display module is unplugged, control returns to the main display module.
NOTE
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
The remote display option cannot be used with the Model 2306-VS
1-4). This remote display module includes all front panel instrument controls/features
When using the remote display, VFD BRIGHTNESS may not appear in the main menu
(dependent on the firmware revision in the unit).
Figure
REMOTE PANEL
ENABLED
When connecting or disconnecting the remote display, allow a few seconds for the
power supply to recognize the action. Fast, repeated connects/disconnects of the
remote display may cause the power supply to hang or appear to hang. Disconnecting
the remote display and waiting a few seconds to reconnect it may clear the problem.
If not, cycling power on the power supply clears the condition.