Series 2230G Triple-Channel Power Supplies User's Manual
Section 3: Function and features

Remote sense connections

To use the remote sense terminals:
1. On the Series 2230G, remove the connections between the positive (+) and sense positive (S+)
terminals.
2. Remove the connection between the negative (–) and sense negative (S-) terminals.
3. Connect the S+ terminal to the positive terminal of the instrument under test.
4. Connect the S- terminal to the negative terminal of the instrument under test.
Figure 12: Remote sense connections to an instrument under test
To ensure the stability of the system, use armored twisted-pair cable between the remote sense
terminals and the instrument under test.
To avoid damage to the Series 2230G, ensure that the polarity of the connections is correct.
2230G-900-01 Rev. A / June 2018
3-19