Keithley 2302 Instruction Manual page 307

Battery/charger simulator
Hide thumbs Also See for 2302:
Table of Contents

Advertisement

Table C-1
Remote calibration command summary
Command
:CALibration
:PROTected
:CODE '<code>'
:COUNt?
:DATE <yyyy>,<mm>,<dd>
:DATE?
:INIT
:SAVE
:LOCK
:STEP0 <nrf>
:STEP1 <nrf>
:STEP2 <nrf>
:STEP3
:STEP4 <nrf>
:STEP5 <nrf>
:STEP6 <nrf>
:STEP7
:STEP8 <nrf>
:STEP9 <nrf>
:STEP10 <nrf>
:STEP11 <nrf>
:STEP12
:STEP13 <nrf>
:STEP14 <nrf>
:STEP15 <nrf>
:STEP16
:STEP17 <nrf>
* Calibration data will not be saved if:
1. Calibration was not unlocked with :CODE command.
2. Invalid data exists. (For example, cal step failed or was aborted.)
3. Incomplete number of cal steps were performed.
4. Calibration was not performed in the proper sequence.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Description
Calibration subsystem.
Cal commands protected by password.
Unlock cal; changes code if cal is already unlocked. (Default
code: KI002306.)
Query number of times 2306 has been calibrated.
Program calibration year, month, day.
Query calibration year, month, day.
Initiate calibration (must be sent before other cal steps).
Save calibration data to EEPROM.*
Lock out calibration. (Abort if calibration is incomplete.)
Output Ch. 1 full-scale voltage (14V).
Calibrate Ch. 1 output voltage using external DMM reading.
Calibrate Ch. 1 voltage measuring using external DMM reading.
Perform Ch. 1 DVM input full-scale (14V) cal.
Output current (1.9A) for Ch. 1 5A full-scale cal.
Calibrate Ch. 1 output current limit using calculated current.
Calibrate Ch. 1 5A measurement range using calculated current.
Output 5mA nominal current for Ch. 1 5mA range full-scale cal.
Calibrate Ch. 1 5mA measurement range.
Output Ch. 2 full-scale voltage (14V).
Calibrate Ch. 2 output voltage setting using DMM reading.
Calibrate Ch. 2 voltage measuring using external DMM reading.
Perform Ch. 2 DVM input full-scale (14V) cal.
Output current (1.9A) for Ch. 2 5A full-scale cal.
Calibrate Ch. 2 output current limit using calculated current.
Calibrate Ch. 2 5A measurement range using calculated current.
Output 5mA nominal current for Ch. 2 5mA range full-scale cal.
Calibrate Ch. 2 5mA measurement range.
Calibration Reference C-3