When the MEASure? command is sent, the specified function is selected and then the
READ? is executed. When the MEASure:ARRay? command is sent, the specified function is
selected and the READ:ARRay? command is executed. See READ? and READ:ARRay? for
details.
If a function is not specified, the measurement(s) will be performed on the active channel's
function that is presently selected.
NOTE
There are no AVERage commands for long integration measurements. The array size
for long integration readings is fixed at one. Therefore, MEASure:LINTegration? and
MEASure:ARRay:LINTegration? are basically the same.
BOTHTRG
BOTHFETCH?
ing
BOTHREAD?
ings
When sending either the BOTHTRG or the BOTHREAD? command, make note that the
command is applied to channel 1 (battery channel) first and then to channel 2 (charger channel).
This allows both channels' triggers to be controlled with a single bus command.
The BOTHTRG, BOTHFETCH?, and BOTHREAD? commands work with the
Models 2306, 2306-VS, and 2306-PJ, but do not work with the Models 2302 or 2302-PJ (i.e.,
the commands do not work with single channel models).