3.
4.
5.
6.
7.
8.
Figure 14-2
Connections for 5A/500mA current calibration
PREV
DCV
ACV
DCI
DISPLAY
NEXT
REL
TRIG
STORE RECALL
POWER
INFO
LOCAL
CHAN SCAN
Note: Use 4-wire connections
to resistor terminals.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
The unit will prompt you for the DMM reading, which will be used to calibrate the
Channel #1 full-scale output voltage:
CAL VOLTS CHAN 1
READ1 14.0000 V
Using the edit keys, adjust the Model 2306 voltage display to agree with the DMM
voltage reading, then press ENTER. The unit will then prompt for another DMM
reading, which will be used to calibrate the full-scale Channel #1 measurement function:
CAL VOLTS CHAN 1
READ2 14.0000 V
Using the edit keys, adjust the display to agree with the new DMM voltage reading, then
press ENTER. The unit will prompt for Channel #1 DVM full-scale calibration:
CAL DVM CHAN 1
ALL READY TO DO?
Press ENTER to complete Channel #1 DVM full-scale calibration.
Connect the digital multimeter volts input and characterized 4Ω resistor to the
Model 2306 OUTPUT #1 SOURCE terminals, as shown in
observe proper polarity (SOURCE + to DMM INPUT HI; SOURCE - to INPUT LO).
Be sure the digital multimeter DC volts function and auto-ranging are still selected.
2001 MULTIMETER
Ω2
Ω4
ACI
FREQ
TEMP
RANGE
AUTO
FILTER MATH
RANGE
CONFIG MENU
EXIT
ENTER
Model 2001 DMM
Input HI
Input LO
SENSE
INPUT
Ω 4 WIRE
HI
350V
1100V
!
PEAK
PEAK
LO
500V
PEAK
INPUTS
F
R
FRONT/REAR
2A 250V
AMPS
CAL
Sense +
Source +
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
OUTPUT #1
SOURCE
+
+
CAT
RELAY
CONTROL
24VDC MAX
FROM FRONT PANEL MENU)
MADE IN
U.S.A.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Model 2306
Calibration
Figure
14-2. Be sure to
OUTPUT #1 connections
shown. Use OUTPUT #2
for Charger Channel.
4 Ω Resistor
Sense -
Source -
DVM IN
OUTPUT #2
+30 VDC MAX
SENSE
SOURCE
DVM
IN
SOURCE
SENSE
SOURCE
DVM
IN
+ _
_
_
_
+ _
_
_
_
+
+
+
+
ISOLATION FROM EARTH: 22 VOLTS MAX.
I