Model 2306 measures them first.) Also set TLEV1 for the initial step. This value needs to be
appropriate for detecting the first DOWN step as an UP step measurement (in
value is set at 750mA). For the UP steps, set the trigger level to a value appropriate for each
rising step. The key to detecting this pulse sequence is setting the step timeout to a value high
enough to bypass the remaining down steps after measuring the first step.
For
Expected measurements
UP
DOWN
TLEVs (all rising)
This pulse sequence can be measured using the continuous pulse method (see the
programming example
be measured using the one-shot method. For the one-shot method, the first step trigger level
value could be any value for detecting the 900 milliamp step.

Timeout setting

TOUT (TimeOUT — timeout setting) specifies the timeout length for detecting a given pulse
step. When the TOUT value is reached, an overflow value for that step reading is returned.
Although all step measurements after the first TOUT step are returned as overflow readings, all
step measurements performed before TOUT was exceeded will have correct readings.
Two timeout settings are used: one for the initial step and another for the rest of the active
steps. The setting for the initial timeout should be set slightly longer than the period of the pulse
for continuous pulse trains. The other timeout setting should cover the longest step duration.
Also, make sure to account for trigger delays when determining timeout settings. There are two
possible trigger delays: the internal trigger delay (15µsec necessary for code execution), and any
user specified trigger delay (optional). The trigger delays occur before the integration process
begins but after pulse detection.
To use the pulse current step method to measure a one-shot pulse train, set the initial timeout
to the maximum setting of 60 seconds. This allows the Model 2306 to be triggered for step
measurements, then a few seconds of delay before generating the one-shot pulse train. The few
seconds of delay are required to ensure the Model 2306 is setup and ready to detect the first step
when it happens along with the rest of the steps.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Figure
3-9, the following expected measurement values and TLEVs were used:
450mA, 575mA, and 700mA (4th–6th pulses)
900mA, 600mA, 300mA (1st-3rd pulses)
TLEV1
750mA (1st pulse)
TLEV2
400mA (4th pulse)
"Continuous pulse train" on page
TLEV3
500mA (5th pulse)
TLEV4
625mA (6th pulse)
3-34). Similarly, this pulse train could
Figure
3-9, this

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2302-pj23062306-vs2306-pj

Table of Contents