Event Enable Registers - Keithley 2302 Instruction Manual

Battery/charger simulator
Hide thumbs Also See for 2302:
Table of Contents

Advertisement

8-18
Status Structure

Event enable registers

As
Figure 8-1
logically ANDed (&) to a corresponding enable bit of an enable register. Therefore, when an
event bit is set and the corresponding enable bit is set (as programmed by the user), the output
(summary) of the register will set to 1, which in turn sets the summary bit of the status byte
register.
The commands to program and read the event enable registers are listed in
details on programming and reading registers,
"Reading registers" on page
NOTE
Table 8-5
Common and SCPI commands — event enable registers
Command
*ESE <NRf>
*ESE?
STATus
:OPERation
:ENABle <NRf>
:ENABle?
:MEASurement
:ENABle <NRf>
:ENABle?
:QUEStionable
:ENABle <NRf>
:ENABle?
Parameters:
<NRf> = 0 to 65535
Note: Power-up and STATus:PRESet resets all bits of all enable registers to 0. *CLS has no effect.
STATus:PRESet has no effect on settings for *ESE.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
shows, each status register set has an enable register. Each event register bit is
8-5.
The bits of any enable register can be reset to 0 by sending the 0 parameter value with
the appropriate enable command (i.e. STATus:OPERation:ENABle 0).
Description
Program standard event enable register (see "Parameters").
Read standard event enable register.
STATus subsystem:
Operation event enable register:
Program enable register (see "Parameters").
Read enable register.
Measurement event enable register:
Program enable register (see "Parameters").
Read enable register.
Questionable event enable register:
Program enable register (see "Parameters").
Read enable register:
Decimal format
see "Programming enable registers"
Table
8-5. For
and
Default
(Note)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2302-pj23062306-vs2306-pj

Table of Contents