Keithley 2302 Instruction Manual page 237

Battery/charger simulator
Hide thumbs Also See for 2302:
Table of Contents

Advertisement

1.
Select the DMM DC volts function, and enable auto-ranging.
2.
Select Channel #1 by pressing the
Channel #2, and enable the Model 2306 DVM input.
3.
Turn on the Model 2306 source output.
4.
Verify digital voltmeter input accuracy for each of the voltages listed in
each test point:
• Set the voltage to the indicated value as measured by the digital multimeter.
• Set current compliance to 0.25A.
• Allow the reading to settle.
• Verify that the Model 2306 voltage reading is within the limits given in the table.
1.
Repeat steps 1 through 5 for Channel #2 (use the
Channel #1 and Channel #2).
Table 13-8
Digital voltmeter input accuracy limits
Model 2306 output
1
Voltage setting
5.00V
10.00V
14.00V
1
As measured by digital multimeter.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
and
keys to toggle between Channel #1 and
Digital voltmeter input reading limits
(2 Years, 18°C-28°C)
4.995 to 5.005V
9.992 to 10.008V
13.990 to 14.010V
Performance Verification
Table
and
keys to toggle between
13-17