Figure 14-3
Connections for 5mA range calibration
PREV
DCV
ACV
DCI
DISPLAY
NEXT
REL
TRIG
STORE RECALL
POWER
INFO
LOCAL
CHAN SCAN
Note: Use 4-wire connections
to resistor terminals.
Step 3: Perform Channel #2 (charger channel) calibration steps
1.
2.
3.
4.
5.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
2001 MULTIMETER
Ω2
Ω4
ACI
FREQ
TEMP
RANGE
AUTO
FILTER MATH
RANGE
CONFIG MENU
EXIT
ENTER
Model 2001 DMM
Connect the OUTPUT #2 SOURCE, SENSE, and DVM IN terminals to the DMM. Use
the same general connections shown in
OUTPUT #2 terminals.
DMM input connections:
HI: SOURCE +, SENSE +, and DVM IN +
LO: SOURCE -, SENSE -, and DVM IN -
The Model 2306 will prompt you to set the full-scale Channel #2 output voltage:
CAL VOLTS CHAN 2
SET 14.0000 V
Use the edit keys to set the voltage to 14.0000V, then press ENTER.
The unit will prompt you for the DMM reading, which will be used to calibrate the
Channel #2 full-scale output voltage:
CAL VOLTS CHAN 2
READ1 14.0000 V
Using the edit keys, adjust the Model 2306 voltage display to agree with the DMM volt-
age reading, then press ENTER. The unit will prompt for another DMM reading, which
will be used to calibrate the full-scale Channel #1 measurement function:
CAL VOLTS CHAN 2
READ2 14.0000 V
Input HI
SENSE
Ω 4 WIRE
INPUT
Input LO
HI
350V
1100V
!
PEAK
PEAK
LO
500V
PEAK
INPUTS
F
R
FRONT/REAR
2A 250V
AMPS
CAL
Sense +
Source +
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
OUTPUT #1
SOURCE
SENSE
+ _
+
+
CAT
I
RELAY
CONTROL
24VDC MAX
(ENTER IEEE ADDRESS
FROM FRONT PANEL MENU)
MADE IN
U.S.A.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
Model 2306
Figure 14-1
except make all connections to the
Calibration
OUTPUT #1 connections
shown. Use OUTPUT #2
for Charger Channel.
3k Ω Resistor
Sense -
Source -
DVM IN
OUTPUT #2
+30 VDC MAX
DVM
IN
SOURCE
DVM
IN
SOURCE
SENSE
SOURCE
_
_
_
+ _
_
_
_
+
+
+
+
ISOLATION FROM EARTH: 22 VOLTS MAX.
LINE FUSE
SLOWBLOW
2.0A, 250V
LINE RATING
100-120VAC, 200-240VAC
50, 60 HZ
165VA MAX
IEEE-488
REMOTE
DISPLAY
OPTION
14-9