Keithley 2302 Instruction Manual page 243

Battery/charger simulator
Hide thumbs Also See for 2302:
Table of Contents

Advertisement

4.
5.
6.
7.
Table 14-2
Model 2306 front panel calibration summary
1
Steps
Description
0, 9
Output 14V
1, 10
Full-scale output voltage
2, 11
Full-scale measure
3, 12
Full-scale DVM
4, 13
5A range output current
5, 14
5A current limit
6, 15
5A measure
7
Model 2306/2306-VS: 5mA range output current
Model 2306-PJ: 500mA range output current
8
Model 2306/2306-VS: 5mA measure
Model 2306-PJ: 500mA measure
16
5mA range output current
17
5mA measure
1
Step numbers correspond to :CAL:PROT:STEP command numbers. See
Steps 0 - 8 calibrate Channel #1 (Battery Channel). Steps 9 - 17 calibrate Channel #2 (Charger Channel).
2
Factory default display values shown.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Press the
key. The unit will then prompt you to run calibration:
CALIBRATE UNIT
RUN
Press ENTER. The unit will then prompt for the calibration code:
CALIBRATE UNIT
Cal Code KI002306
Using the edit keys, set the display to the current calibration code, then press ENTER.
(Default: KI002306.) The unit will then prompt you as to whether or not to change the
code:
CALIBRATE UNIT
Change Code NO
Be sure NO is selected (use the
below to calibrate the unit. (See Changing the calibration code at the end of this section
if you wish to change the code.)
and
keys), press ENTER, then follow the steps
Nominal calibration
2
signal
14V
Figure 14-1 on page 14-6
14V
Figure 14-1 on page 14-6
14V
Figure 14-1 on page 14-6
14V
Figure 14-1 on page 14-6
1.9A
Figure 14-2 on page 14-7
1.9A
Figure 14-2 on page 14-7
1.9A
Figure 14-2 on page 14-7
4.5mA
450mA
4.5mA
450mA
4.5mA
Figure 14-3 on page 14-9
4.5mA
Figure 14-3 on page 14-9
Table
14-3.
Calibration
14-5
Test connections