Keithley 2302 Instruction Manual page 206

Battery/charger simulator
Hide thumbs Also See for 2302:
Table of Contents

Advertisement

12-8
SCPI Tables
Table 12-4 (cont.)
SENSe command summary (refer to Tables 2-3, 3-2, and 4-2) (cont.)
Command
:SENSe[1]
:PCURrent
:STEP
:DELay
<NRf>
:SEQuence
:LENgth
:SKIP
:DELay?
:RANGe
<NRf>
:RANGe?
:MILLiamp
<NRf>
:MILLiamp?
:TLEVx
<NRf>
:TLEVx?
:SYNChronize
[:STATe] <b>
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Description
Step delay: 0msec–100msec (in 10μsec steps).
Define up to 19 delay sequence values using the index
and the time value (e.g. "1,1e-3" sets the first delay
to 1ms). Ranges: i: 1-19 and t: 0 to 100e-3.
Define the delay sequence length (0 to 19). Specify
zero to always use the PCUR:STEP:DELay value.
Specify whether (1) or not (0) to skip sequential
triggers.
1=Only requires 1 trigger
0 = Requires multiple triggers
Query setting for step delay.
Set step trigger level range for 5A current range:
100mA, 1A, 5A.
Query setting for step trigger level range for 5A
current range.
Model 2306-PJ only — when on 500mA current
range. The parameter <NRf> sent with this
command causes the trigger to be set with the
trigger level setting of HALFamp, HUNDred, or
TEN.
Model 2306-PJ only — Query the trigger level
setting on the 500mA current range. Receives
responses in amps of 0.5, 0.1, or 0.01,
accordingly.
Set trigger level for each TLEV step (x equals 1–20)
(0.0–maxA where max is 100mA for 100mA
range, 1A for 1A range, and 5A for 5A range
setting).
Query setting for trigger level for each TLEV step
(x equals 1–20).
Pulse detection triggering:
Send ON to select pulse current measurements or OFF
to select pulse current digitization.
Default
SCPI
parameter
0
0
0
5A
0.0
ON