Keithley 2302 Instruction Manual page 227

Battery/charger simulator
Hide thumbs Also See for 2302:
Table of Contents

Advertisement

2.
Select the multimeter DC volts measuring function, and enable auto-ranging.
3.
Select Channel #1 by pressing the
Channel #2. Make sure the Model 2306 output is turned on.
4.
Verify output voltage accuracy for each of the voltages listed in
point:
• Set the Model 2306 output voltage to the indicated value.
• A value of current compliance of 0.25A is sufficient.
• Allow the reading to settle.
• Verify that the multimeter reading is within the limits given in the table.
1.
Repeat steps 1 through 4 for Channel #2 (use the
Channel #1 and Channel #2).
Table 13-2
Output voltage accuracy limits
Model 2306 output
voltage setting
5.00V
10.00V
15.00V
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
and
keys to toggle between Channel #1 and
Output voltage limits
(2 Years, 18°C-28°C)
4.9945 to 5.0055V
9.992 to 10.008V
14.9895 to 15.0105V
Performance Verification
Table
13-2. For each test
and
keys to toggle between
13-7

Hide quick links: