Changing the calibration code

The default calibration code may be changed from the front panel or via remote as discussed
below.

Changing the code from the front panel

Follow the steps below to change the code from the front panel:
1.
2.
3.
4.
5.
6.
7.
8.

Changing the code by remote

To change the calibration code by remote, first send the present code, then send the new code.
For example, the following command sequence changes the code from the 'KI002306' remote
default to 'KI_CAL':
Use any combination of letters and numbers up to a maximum of eight characters.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Press the MENU key, then choose CALIBRATE UNIT, and press ENTER. The
instrument will display the last date calibrated:
CALIBRATE UNIT
LAST ON 02/01/2003
Press the
key. The instrument will display the number of times it was calibrated:
CALIBRATE UNIT
TIMES = 1
Press the
key. The unit will then prompt you to run calibration:
CALIBRATE UNIT
RUN
Press ENTER. The unit will then prompt for the calibration code:
CALIBRATE UNIT
Cal Code KI002306
Using the edit keys, set the display to the current present calibration code then press
ENTER. (Default: KI002306.) The unit will prompt you as to whether or not to change
the code:
CALIBRATE UNIT
Change Code NO
Select YES then press ENTER. The instrument will prompt you to change the code:
CALIBRATE UNIT
New Code: KI002306
Use the edit keys to set the new code then press ENTER to accept the new code.
Press the MENU key to exit calibration and return to the main menu.
:CAL:PROT:CODE 'KI002306'
:CAL:PROT:CODE 'KI_CAL'