F-2
Model 2302 Specifics
General information
NOTE
Specifications
Full power supply specifications can be found in Appendix A of this manual.
Power supply overview
The Model 2302 and 2302-PJ power supplies (single channel battery simulator —
Figure
documented in this manual (2306-901-01).
Operational differences
Front panel operation
Since the Model 2302 and 2302-PJ are single channel units, disregard any statements in the
manual such as "Toggle active channel using the
active channel is not displayed—the active channel is always the battery channel.
SCPI operation
If a channel 2 bus command is issued, an "Undefined header" error (-113) will result.
*IDN? will reflect the Model 2302 and 2302-PJ instead of the Model 2306, 2306-VS, or
2306-PJ.
When using the DISP:CHAN command, only send a "1" as a parameter. Other values will
result in a parameter data out of range error (-222).
The BOTHOUTON and BOTHOUTOFF along with other BOTH commands do not work
with the Model 2302 or 2302-PJ.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Since the Model 2302 and 2302-PJ are single channel battery simulators, functions
contained in this manual related to the second channel (i.e., the charger channel) are
not available for the Model 2302 and 2302-PJ. With this in mind, for the Model 2302
and 2302-PJ, only battery channel features contained in this manual apply.
F-1) simulate a battery. The features match the battery channel features (Channel 1)
or
keys." Also, the #1 or #2 indicating the