Using Fast, Search, And Detect - Keithley 2302 Instruction Manual

Battery/charger simulator
Hide thumbs Also See for 2302:
Table of Contents

Advertisement

•OFF or 0 - Disables trigger synchronization and selects pulse current digitization. See
"Pulse current digitization" for details on digitizing a current pulse or waveform.
:TLEVel Commands
A valid trigger level for detecting the pulse is needed whether trigger synchronization is
ON or OFF (see :SYNChronize commands above).
SENSe[1]:PCURrent:SYNChronize:DELay <NRf>
SENSe2:PCURrent:SYNChronize:DELay <NRf>
1. The smallest step size for trigger delay is 10µsec. If you specify a smaller step size, it is
adjusted up to the next 10µsec step value (e.g., 43µsec is adjusted up to 50µsec).
2. After pulse detection but before the integration process begins, the internal trigger delay
of 15µsec (for code execution) in addition to any user specified trigger delay must elapse.
This command is used to set the user trigger delay.
3. Make sure this setting works with the :TIME settings to produce an accurate reading.
Although AUTO accounts for internal trigger delay, HIGH, LOW, and AVERage do not.
Note that none of the TIME commands account for the user trigger delay.
SENSe[1]: PulseCURrent:FAST <b>
SENSe2: PulseCURrent:FAST <b>
Refer to "Using FAST, SEARch, and DETect" for detailed usage information.
SENSe[1]:PulseCURrent:SEARch <b>
SENSe2: PulseCURrent:SEARch <b>
Refer to "Using FAST, SEARch, and DETect" for detailed usage information.
SENSe[1]: PulseCURrent:DETect <b>
SENSe2: PulseCURrent:DETect <b>
Refer to "Using FAST, SEARch, and DETect" for detailed usage information.
SENSe[1]:PulseCURrent:TimeOUT <NRf>
SENSe2: PulseCURrent: TimeOUT <NRf>
Refer to "Using FAST, SEARch, and DETect" for detailed usage information on properly
setting the TimeOUT variable.

Using FAST, SEARch, and DETect

Use FAST, SEARch, and DETect to control how background readings are taken. A
background reading is a measurement taken by the power supply between user triggered
readings. The selected function dictates how background readings are taken between user
triggered readings.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Pulse Current Measurements
Applies to battery channel (#1)
Applies to charger channel (#2)
Applies to battery channel (#1)
Applies to charger channel (#2)
Applies to battery channel (#1)
Applies to charger channel (#2)
Applies to battery channel (#1)
Applies to charger channel (#2)
Applies to battery channel (#1)
Applies to charger channel (#2)
3-17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2302-pj23062306-vs2306-pj