Example 2. Voltage step and measure high current in a chain of GSM signal
Model 2306-VS configuration using channel 1:
*RST
:VOLT 6
:CURR 2.25
:SENS:PCUR:FAST ON
:SENS:PCUR:TIME:HIGH 500e-6
:SENS:PCUR:MODE HIGH
:SENS:FUNC 'PCUR'
:TRIG:EXT:STEP:VOLT ON
:TRIG:EXT:STEP:READ AUTO
:TRIG:EXT:STEP 1,5.7,0
:TRIG:EXT:STEP 2,5.9,0
:TRIG:EXT:STEP 3,6.1,0
:TRIG:EXT:STEP 4,6.3,0
:TRIG:EXT:STEP:POIN 4
Loop1:
:TRIG:EXT:ENAB ON
monitor for the trigger out on channel 1
Loop2:
generate 4 trigger in pulses on channel 1 and look for trigger out pulses on
channel 1 - make sure a new trigger in pulse occurs after each trigger out
pulse. Trigger in pulses are ignored until the trigger out pulses. Trigger
ins detected on falling edge and trigger outs pulse low for about 10us.
:FETCH:ARRAY?
:TRIG:EXT:ENAB:INIT
go back to Loop 2
:TRIG:EXT:ENAB OFF
go back to Loop 1 after making changes
NOTE
If you are unable to monitor for trigger out pulses, either:
(1) monitor the buffer full bit in the status model (Section 8), or
(2) allow sufficient time between trigger in pulses to allow the unit to complete each
step (see the specifications in Appendix A for time guidelines).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
External Triggering (Model 2306-VS Only)
' reset the unit
' set voltage
' set current
' turn fast mode on
' set high current time (this must be set
' in external triggering mode because
' no AUTO detection exists)
' measure high current
' switch to pulse current functionality
' turn ON the voltage step
' switch read mode to AUTO, means
' 2306-VS will measure right after it is
' triggered
' step 1 is 5.7 volts with 0 delay
' step 2 is 5.9 volts with 0 delay
' step 3 is 6.1 volts with 0 delay
' step 4 is 6.3 volts with 0 delay
' set the number of points to be
' triggered and put into the buffer after
' voltage stepping.
' turn external triggering ON. 2306-VS
' is waiting for the trigger signal.
' get the data back
' init external trigger to step and take
' another set of 4 measurements
' when done stepping and taking
' measurements or desire to make
' changes
6-17