Current Measurement Accuracy Test (Channel 1); Current Measurement Accuracy Test (Channel 2) - Agilent Technologies 4349B Manual

4-channel high resistance meter
Hide thumbs Also See for 4349B:
Table of Contents

Advertisement

4349B

Current Measurement Accuracy Test (Channel 1)

Test
Resistor
Voltage
Current
Setting
Setting
100 pA
10
10
100 pA
10
10
100 pA
10
10
1 nA
10
9
1 nA
10
9
10 nA
10
8
100 nA
10
7
1  A
10
6
10  A
10
5
100  A
10
5
1 k is the input impedance of the 4349B's ammeter.
1

Current Measurement Accuracy Test (Channel 2)

Test
Resistor
Voltage
Current
Setting
Setting
100 pA
10
10
100 pA
10
10
100 pA
10
10
1 nA
10
9
1 nA
10
9
10 nA
10
8
100 nA
10
7
1  A
10
6
10  A
10
5
100  A
10
5
1 k is the input impedance of the 4349B's ammeter.
1
Measurement
Time
1 V
30 ms
1 V
100 ms
1 V
400 ms
1 V
400 ms
1 V
30 ms
1 V
30 ms
1 V
30 ms
1 V
30 ms
1 V
30 ms
10 V
10 ms
Measurement
Time
1 V
30 ms
1 V
100 ms
1 V
400 ms
1 V
400 ms
1 V
30 ms
1 V
30 ms
1 V
30 ms
1 V
30 ms
1 V
30 ms
10 V
10 ms
Multimeter
Reading [a]
Reading [b]
V
V
V
V
V
V
V
V
V
V
Multimeter
Reading [a]
Reading [b]
V
V
V
V
V
V
V
V
V
V
4349B
Test Result
Equation
pA
b
pA
b
pA
b
nA
b
nA
b
nA
b
nA
b
 A
b
4
 A b
a
4
 A b
a
4349B
Test Result
Equation
pA
b
pA
b
pA
b
nA
b
nA
b
nA
b
nA
b
 A
b
4
 A b
a
4
 A b
a
Maintenance
1
4
a
cv2
4
a
cv2
4
a
cv2
4
a
cv3
4
a
cv3
4
a
cv4
4
a
cv5
4
a
cv6
(cv7+1 k )
(cv7+1 k )
1
4
a
cv2
4
a
cv2
4
a
cv2
4
a
cv3
4
a
cv3
4
a
cv4
4
a
cv5
4
a
cv6
(cv7+1 k )
(cv7+1 k )
9-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents