Working in the Cabinet
Working in the Cell Logic Type A2 biosafety cabinet requires preparation, planning, and
an understanding of aseptic technique to work safely. The steps below represent the
minimum requirements before, during, and after working in a biosafety cabinet. Consult
your Health & Safety department for additional SOP requirements.
Planning
1. Thoroughly understand procedures and equipment required before beginning
work.
2. Arrange for minimal disruptions, such as room traffic or entry into the room while
the cabinet is in use.
Cabinet Start-up
1. Turn off UV light (if included on your cabinet).
2. Slowly raise the sash until the bottom of the sash aligns with the sash indicator
decal located on the left side of the work area.
3. Turn on the light and cabinet blower if the automatic features have not been
enabled.
4. Check the air grilles for obstructions.
5. Allow the cabinet to operate until the
from 'WAIT' to 'OK'. Reference Figure 7-5.
6. Wash hands and arms thoroughly with germicidal soap.
7. Wear appropriate personnel protective equipment (PPE).
Airflow Status
on the
71
Home Screen
updates
Figure 7-5